Yokogawa In-Situ Gas Analyzer TDLS8000

Yokogawa In-Situ Gas Analyzer TDLS8000 product thumbnail
Get in Touch
button-arrow-red Back

Yokogawa’s new TDLS8000 houses all of the industry’s leading features in one robust device. The platform design is for in situ measurements which negate the need for sample extraction and conditioning. The non-contacting sensor allows for a variety of process types including corrosive, abrasive and condensing. The first generation platform has been proven in many others for the measurements of O2, CO, CH4, NH3, H2O and many more NIR absorbing gases. This second generation platform has improved reliability and ease of installation and maintenance while still meeting or exceeding designed application demands.

Details

The most trusted laser analyzer designed specifically to meet all of your requirements in one robust device that is easy to operate and maintain.

Features

  • SIL2 TruePeak combined with smart laser Technology
  • Intuitive touchscreen HMI
  • HART and Modbus TCP communications standard
  • 8-stage auto-gain adapts to difficult applications
  • Fully field repairable with 50 days of data and spectra storage
  • Compact design for one-man installation without sacrificing ruggedness
  • Area classification Zone2/Div2 or Zone1/Div1

System Configuration

High Reliability

  • Reference Cell
    Internal reference cell in the laser module ensures peak locking during trace measurement.
  • Auto gain
    Auto-gain enables wide signal ranges against dynamic variation of transmission.

  • Validation
    Validation can be initiated manually, remotely, or automatically on a daily, weekly or monthly basis defined by the user.
  • SIL2 certified
    IEC61508 SIL designed & approved, SIL2 capability for single analyzer use, SIL3 capability for dual analyzer use.

TruePeak

The TruePeak we can measure the area of the absorbance peak. This eliminates effects from changing background gases, allowing for simple pressure and temperature compensation.

Intuitive touchscreen HMI

  • Touch screen 7.5 inch color LCD on HMI
    Makes it simple to operate
    Gives all the information including trend graph and eliminate PC to maintenance
    Can be remotely installed
    TDLS8000 TouchScreen
  • Mini Display
    Optical transmission at both the ends for easy Alignment
    TDLS8000 display

  • Multilingual capability
  • Heating, cooling and air conditioning
  • Built-in density and enthalpy tables
  • Meter-bus interface and RS485 communications
  • Metric or US units
  • 4-wire Pt100 RTD temperature inputs
  • Fully programmable
  • Calculates and displays Volume, Energy, Temperatures and Peak Energy
  • Watertight to IP66 (Nema 4X)
  • High accuracy

Standard Specifications

Measurement objectO2, CO, CO or CH4, CO2, CO + CO2, H2O, NH3, NH3 + H2O, H2S, HCl concentration in combustion exhaust gas and process gas
Measurement systemTunable diode laser spectroscopy
Measured components and rangeMeasured componentMin. rangeMax. range
O20-1%0-25%
CO(ppm)0-200 ppm0-10,000 ppm
CO+CH4CO0-200 ppm0-10,000 ppm
CH40-5 %
NH30-30 ppm0-50,000 ppm
H2O (ppm) in non HC0-30 ppm0-30,000 ppm
H2O (ppm) in HC0-30 ppm0-30,000 ppm
CO (%)0-20 %0-50 %
CO (%) + CO2 (%)0-30 %0-100 %
 NH3 + H2ONH30-30 ppm0-5,000 ppm
H2O0-5 %0-50 %
H2S0-5 %0-100 %
CO2 (%) High Range0-1 %0-5 %
CO2 (%) Extend.Range0-30 %0-50 %
H2O (%)0-10 %0-100 %
HCl0-50 ppm0-5,000 ppm
Output path lengthOptical distance between the laser unit and the sensor control unit Standard; 0.5 to 6 m, Max; 30 m
Output signal2 points, 4 to 20 mA DC
Output types; Gas concentration, Transmission, Process gas temperature, Process gas pressure
Output range; 3.0 to 21.6 mA DC
Digital communicationHART, Ethernet
Contact outputs2 points, contact rating 24 V DC, 1 A
DO;
Function: Activate during Warning / Calibration / Validation / Warm up / Maintenance conditions
Fault;
Function: Activate during normal condition, not activate during fault condition or when the system power is off
Valve control output2 points
Function; Activate calibration or validation solenoid valves for zero, span or validation gas
Output signal; 24 V DC, 500 mA Max. per terminal
AlarmWarning;
Gas concentration low/high, Transmission low, Process pressure low/high, Process temperature low/high, Validation required, Validation failure, Zero/Span calibration error, Non process alarm, External alarm
Fault;
Laser module temperature low/high, Laser temperature low/high, Detector signal high, Peak center out of range, Reference peak height low, Absorption too high, Transmission lost, Reference transmission low, Reference peak height high, Laser unit failure, Laser module error, File access error, E2PROM access error.
Contact inputs
(Digital input)
2 points
Function;
External alarm/Calibration start/Validation start/Stream switch
Contact specification;
Zero voltage contact input
Input signal; Open signal;
100 kΩ or more, Close signal; 200 Ω or less
Input signal
(Analog input)
2 points, 4 to 24 mA DC
Input types; Process gas temperature, Process gas pressure
Self-diagnosticsLaser Unit temperature, Sensor Control Unit temperature, Laser temperature, Detector signal level, Memory read/write function, Peak locking condition
CalibrationCalibration method; Zero/Span calibration
Calibration mode; Manual, Auto, Local initiate (HMI)
ValidationValidation method; Up to 2 points
Validation mode; Manual, Auto, Local initiate (HMI)
Power supply24 V DC ±10%
Warm-up time5 min.
Protection degreeIP66, NEMA Type 4X
Hazardous area classificationsDivision 1, Zone 1; Explosion-proof/ Flame-proof type; FM, cFM, ATEX, IECEx (Pending)
Division 2, Zone 2; Non-Incendive/Type n; FM, cFM, ATEX, IECEx
Process gas conditionProcess gas temperature; Maximum 1500℃
Process gas pressure; Max. 1 MPa, Min. 90 kPa
Dust in process gas; 20 g/m3 or less
Installation conditionAmbient operating temperature; -20 to 55℃
Storage temperature; -30 to 70℃
Humidity; 0 to 95%RH at 40℃ (Non-condensing)
Mounting flange type; ASME B 16.5, DIN, JIS
Gas connections; 1/4 NPT or Rc1/4
Purge gas;
Recommended purge gasses
O2 analyzer; N2 (99.99% or greater, application dependent)
H2O ppm analyzer; N2 (99.99% or greater with < 20 ppm H2O for feed to the optional dryer package)
CO, CO or CH4, CO2, CO+CO2, NH3, NH3+H2O, H2S, HCl;
N2 (99.99% or greater, application dependent) or Instrument air  CO, NH3 analyzer: N2 (99.99% or greater, application dependent) or Instrument air
Purge gas flow rates;
5-20 L/min for optic
5-30 L/min for process window
Purge gas connections;
1/4NPT (-G1, -C2, -D2, -C2, -D1, -C1),
Rc1/4 (-G2, -S2, -E2, -J2, -E1, -J1)

Performance

Measurement gasRepeatabilityLinearity
O2± 1% reading or ± 0.01% O2, whichever is greater± 1% F.S.
CO (ppm)± 2% reading or ± 1 ppm CO, whichever is greater± 1% F.S.
CO + CH4CO± 1% reading or ± 1 ppm CO, whichever is greater± 2% F.S.
CH4± 4% reading or ± 0.02% CH4, whichever is greater± 4% F.S.
NH3± 2% reading or ± 1 ppm NH3, whichever is greater± 2% F.S.
H2O (ppm) in non
HC
± 2% reading or ± 0.1 ppm H2O, whichever is greater± 1% F.S.
H2O (ppm) in HC± 2% reading or ± 0.1 ppm H2O, whichever is greater± 1% F.S.
CO (%)± 1% reading or ± 0.01% CO, whichever is greater± 1% F.S.
CO (%) + CO2 (%)CO± 1% reading or ± 0.1% CO, whichever is greater± 1% F.S.
CO2± 1% reading or ± 0.1% CO2, whichever is greater± 1% F.S.
NH3 + H2ONH3± 2% reading or ± 1 ppm NH3, whichever is greater± 2% F.S.
H2O± 4% reading or ± 0.05% H2O, whichever is greater± 2% F.S.
H2S± 1% reading or ± 0.005% H2S, whichever is greater± 1% F.S.
CO2 (%) High Range± 1% reading or ± 0.005% CO2, whichever is greater± 1% F.S.
CO2 (%) Extend. Range± 1% reading or ± 0.02% CO2, whichever is greater± 1% F.S.
H2O (%)± 1% reading or ± 0.004% H2O, whichever is greater± 1% F.S.
HCL± 1% reading or ± 2.5 ppm H2O, whichever is greater± 2% F.S.

YH8000 HMI Unit

The YH8000 is a HMI designed specifically for the Tunable Diode Laser Gas Analyzer, the TDLS8000.

  • Touchscreen 7.5 inch color LCD on HMI
  • Makes it simple to operate
  • Can be remotely installed
  • Up to 4 units connection available

Specification

DisplayTouchscreen 7.5 inch TFT color LCD panel, 640 x 480 (VGA)
CommunicationEthernet; RJ-45 connector, Communication speed; 100 Mbps
Protection degree
of enclosure
IP65, NEMA Type 4X
Weight4 kg
MountingAnalyzer mount (Front, left-side, right-side) with tilt function, Pipe mount or Panel mount
Cable Entries1/2NPT or M20 x 2
Installation
conditions
Ambient operating temperature; -20 to 55℃
Storage temperature: -30 to 70℃
Humidity: 10 to 90%RH at 40℃ (Noncondensing)
Power Supply24 V DC ±10%
Hazardous area
classifications
Division 2, Zone2: Non-Incendive/Type n; FM, cFM, ATEX, IECEx

Brochures
  •  Tunable Diode Laser Spectrometer TDLS8000 (1.4 MB)
Instruction Manuals
  •  TDLS8000 Tunable Diode Laser Spectrometer (18.0 MB)
  •  TDLS8000, YH8000, IF8000, YC8000 Protection of Environment (Use in China) (182 KB)
  •  TDLS8000 Tunable Diode Laser Spectrometer (18.6 MB)
General Specifications
  •  Model TDLS8000 Tunable Diode Laser Spectrometer
Software
  •  HART DD file – Tunable Diode Laser Spectrometer TDLS8000 [DevRev1]
  •  Yokogawa DTM HART
Technical Information
  •  TDLS8000 Coating Procedure (271 KB)
  •  Oxygen Analyzer Selection Guide (929 KB)
Certificates
  •  SIL Certificate TDLS8000 (385 KB)
Forms
  •  TDLS Application Data Sheet

[hr]

Resources

Achieve Optimal Combustion of Fired Heater Using TDLS to Reduce Environmental Impacts and Increase Manufacturing Efficiency

Directly Measure Oxygen of Process Gas in 5 Seconds or Fewer. An SIL2 Certified Safe Measurement Solution

Real-time O2 Measurement at Garbage Incinerators Improves Combustion Efficiency and Reduction in NOx and CO

Accurate, Maintenance-free Monitoring of H2S Quantities in a Black Liqor Recovery Boiler

Optimizing Combustion Control with the TDLS8000

Measurement of NH3 Concentrations in Stack Flue Gas Using TDLS200

Measurement of O2 Concentrations in Vacuum Distillation Column in Petroleum Refining

Oxygen Measurement in Vapor Recovery

Electrolysis Plant Trace H2O (Moisture) Measurement by the TDLS200

Measurement of O2 Concentrations in Recovery Flue Gas from Converter Furnace by the TDLS200

News

Corporation of America Releases TDLS8000 Tunable Diode Laser Spectrometer

Product Overviews

Yokogawa Canada

Envision a plant where people are watchful and attentive while your business responds to change quickly and efficiently. Now picture an operation that delivers non-stop production while confidently expanding your capabilities into the future. Imagine no further. This is the vision and promise behind Yokogawa’s VigilantPlant®, the clear path to operational excellence. SEE Clearly, KNOW in Advance, ACT with Agility. Field Sensing & Actuation | Production Control | Network-based Control | High-reliability SCADA Safety Management Systems | Intelligent Production Management | Asset Management

EJX630A In-Line Mount High Perfomance Gauge Pressure Transmitter

Have any questions?

We'd love to help

Get in touch

Call Us Toll Free Across Canada!

1.888.388.1592

Corporate Headquarters

101-11127 15 Street NE,
Calgary, AB Canada T3K 2M4
1.403.250.1594
edge-footer-top-blue footer-tower
  • LinkedIn
  • Twitter
  • YouTube
  • Facebook
View all Accreditations
  • COR
  • ISN
  • Avetta
A sillouette of an oil and gas plant.

© 2020 MicroWatt Controls | Website by True Market